10. Reprocessing of reusable medical devices and other devices in general practice

Components of Steriliser Validation


        1. Components of Steriliser Validation

Last revised: 12 Jun 2026

Components of Steriliser Validation

For each sterilisation cycle used – that is, the specific time, temperature, and pressure parameters – the following needs to be included in the validation:

  • Heat distribution testing on an empty chamber (may be sourced from manufacturer or prior validation)
  • Description and diagram of the challenge pack/load and loading configuration
  • Measurement of penetration time for the challenge pack
  • Physical qualification, including confirmation of the time at temperature using calibrated instruments
  • Microbiological qualification, using biological indicators across three identical consecutive cycles.

Table 10.20 Steriliser validation and maintenance summary

Action / Test

Responsibility

Frequency

Purpose / Description

Record required

Full validation (IQ, OQ, PQ)

Qualified technician

On installation/commissioning
After relocation, upgrade, or major repair
Annually
After significant process changes

Confirms consistent steriliser performance with typical loads

Yes – detailed report

Heat distribution test

Manufacturer or qualified technician

At initial validation or as provided

Confirms uniform heat distribution

Yes – validation report

Challenge pack / Load description

Qualified technician

Each annual validation

Document typical load for validation

Yes – included in report

Staff competency check

IPC Coordinator / Practice manager

During validation and annually

Confirm staff understand reprocessing protocols

Yes – training records

Biological indicator testing (PQ phase)

Technician or Reprocessing team

During validation and as needed

Confirms sterility during validation using reference loads

Yes – indicator results

Cycle monitoring – Biological

Reprocessing team

During validation or troubleshooting

Used across 3 cycles to confirm sterility

Yes – keep results

Review of validation report

Reprocessing team

Each time validation is performed

Ensures technician’s results are understood and applied

Yes – review notes

Leak-rate test (vacuum sterilisers) a

Reprocessing team

Daily (or weekly if air detector fitted), or according to manufacturer recommendations or frequency of use.

Checks that the chamber is airtight

Yes – log result

Bowie-Dick / Steam penetration test b

Reprocessing team

Daily (if required according to the type of steriliser is used and what devices are being sterilised)

Confirms air removal and steam penetration

Yes – record result

Cycle monitoring – Mechanical

Reprocessing team

Every cycle

Review auto printout or data logger

Yes – retain printout

Cycle monitoring – Chemical

Reprocessing team

Every cycle

Indicators in every pack or tray

Yes – note checked and passed

Functional test after offsite servicing

Reprocessing team

On return before reuse

Vacuum, Bowie-Dick, and indicator tests

Yes – test outcomes

Enhanced monitoring if validation delayed

Reprocessing team

Temporary

Use high-level indicators in every load

Yes – all indicators logged

Note: Equipment needs to be requalified at least annually and whenever significant changes occur (eg repairs, new processes, new equipment). Requalification may involve full or partial validation (IQ, OQ, PQ), depending on the risk and nature of the change. A competent person needs to assess and document the need for testing, the rationale, outcomes, and any corrective actions.

a For a steriliser that uses a vacuum in any part of its operation, a leak-rate test is required on each day of its use (or weekly, if there is an air detector) to check the leak rate does not exceed the maximum specified in the current relevant standard. Refer to the steriliser operating instructions for information on whether an air detector is fitted and how to perform the leak-rate test.

b If the practice uses Class B sterilisation cycles and reprocessed narrow-lumen (hollow A) items, a steam penetration/air removal test needs to be performed daily in accordance with steriliser instructions for use and process challenge device instructions for use. Steam penetration tests (eg Bowie-Dick) are a form of process challenge device. These simulate worst-case sterilisation conditions, such as narrow lumens, and confirm air removal and steam penetration during vacuum cycles. The helix device is commonly used for small benchtop sterilisers.

Table 10.21 Washer-disinfector validation and maintenance summary

Action / Test

Responsibility

Frequency

Purpose / Description

Record required

Full validation (IQ, OQ, PQ)

Qualified technician

On installation/commissioning
After relocation or repair
Annually

Confirms correct function, cleaning efficacy, and temperature control

Yes – validation report

Water quality testing

Reprocessing team or External

Monthly (or more)

Test for hardness, chloride, conductivity

Yes – record results

Review of validation report

Reprocessing team

Each time validation is performed

Ensure technician’s results are applied in practice

Yes – review checklist

Cycle monitoring – mechanical

Reprocessing team

Every cycle

Check temp, time, chemical dose, pressure, flow

Yes – printout or log

Visual inspection of items

Reprocessing team

Every cycle

Inspect cleaned items for visible debris

Optional – note if recleaning needed

Cleaning efficacy testing (optional)

Reprocessing team

As required (eg complex instruments)

Use soil tests or surrogate devices if validated

Yes – if used

Table 10.22 Ultrasonic cleaner validation and maintenance summary a

Action / Test

Responsibility

Frequency

Purpose / Description

Record required

Validation (functionality and cleaning efficacy testing) b

Qualified technician

On installation/commissioning
After relocation or repair
Annually

Confirm cleaning efficacy and cavitation performance

Yes – validation report

Water quality monitoring

Reprocessing team

Monthly (or more)

Test rinse water quality if applicable

Yes – water test log

Functionality Test

Reprocessing team

Daily

Confirm cleaner is operating correctly per AS 2773

Yes – daily log

Log solution changes and degassing (if applicable)

Reprocessing team

Every cycle

Confirms safe and effective operation of cleaning cycle

Yes – if used

Visual Inspection of Items

Reprocessing team

Every cycle

Check items after ultrasonic cleaning

Optional – note failures

a An ultrasonic cleaner is optional for pre-cleaning some reusable medical devices (it is not required in general practice).

b Ultrasonic cleaners are pre-cleaning devices. They do not disinfect or sterilise, so the validation approach is simpler, but still essential when used in a validated reprocessing pathway.

Table 10.23 Routine cleaning and monitoring summary

Equipment / Process

Responsibility

Frequency

Description

Record required

Manual cleaning

Reprocessing team

Every item, every time

Visual check for cleanliness

Optional – if failed, reprocess

Cleaning efficacy (Was is clean?)

Reprocessing team

Every item

Visually inspect each item after cleaning. Reclean if dirty. Use test devices or soil tests for complex items if available and validated.

Optional – if failed, reprocess

Heat sealer check (if used/applicable)

Reprocessing team

Each session

Check seal function and one test pouch

Yes – note temp & pouch test

Sterilisation containers

Reprocessing team

Before each use

Check for damage, gasket, filters

Optional – in SOP

Packaging checks

Reprocessing team

Every pack

Inspect seal and follow procedure

Optional – redo and document if failed


Note: The above tables summarise typical validation and monitoring requirements for commonly used reprocessing equipment in general practice. If a practice introduces any additional equipment – for example drying cabinets, low temperature sterilisation systems, or other specialised devices – that equipment needs to be:

  • installed and set up according to the manufacturer’s instructions
  • serviced and validated by a qualified technician prior to use
  • included in routine monitoring and recordkeeping schedules.

Validation requirements for such equipment needs to align with relevant standards and manufacturer specifications.

Advertising