Before moving an instrument to the next stage of cleaning, disinfecting, or sterilising, the practice needs to check that the previous stage has been done properly.
A device needs to be clean before it can be disinfected or sterilised.
Always check that the equipment used has run its cycle correctly. This includes reviewing cycle printouts or digital records.
There should be a clear system in place to identify processed (ready for use) and unprocessed (dirty) instruments.
When a device can be released for use
A device can only be released for patient use when it has passed all relevant checks – it is clean, sterilised, and its packaging is intact. Table 10.25 outlines the minimum criteria that need to be met at each phase of reprocessing. A device can only be released for patient use once all criteria across all applicable stages have been met. Skipping or partially completing any stage applicable to a reusable medical device renders the device non-compliant.
Table 10.25 Minimum criteria for releasing reusable medical devices
|
Stage of reprocessing
|
Release criteria
|
|
Manual cleaning
|
Device is visually clean and dry.
|
|
Automated cleaning
|
Device is visually clean and dry and cycle records match process settings.
|
|
Thermal disinfection (heat-based)
|
Device is visually clean and dry, and cycle records confirm the correct temperature and time were reached.
|
|
Packaging – Sealable pouches/reels (PSBS)
|
Packaging is the right size, labelled correctly, intact, and includes a chemical indicator that has changed to the correct colour.
|
|
Moist heat sterilisation (with drying)
|
Cycle records confirm the correct time, temperature, and pressure were reached. External chemical indicators show the correct colour change. Packs are dry and intact. Data-logged information or a printout is checked for every load (an on-screen display alone is not sufficient). Chemical indicators inside the steriliser show the correct result. Process challenge devices, where used, show the expected result.
|
|
⚠️ All listed criteria across all applicable stages need to be met before a device can be released for patient use. If any requirement is not met, the device is considered non-compliant and needs to be quarantined and managed under the formal procedure.
|
If any of the requirements are not met, the device is considered non-compliant.
These non-compliant devices need to be quarantined and dealt with under a formal procedure for handling such issues (this is a quality control measure to prevent use of faulty items).