Correct loading is critical to:
- ensure steam can reach and sterilise every surface of every item,
- protect sterile barrier systems from damage, and
- allow complete drying of the load.
General principles for loading the steriliser
- Only sterilise instruments and devices that are:
- compatible with the steriliser’s specifications, and
- included in the practice’s validated load configuration.
- Do not overload. Each load needs to remain within the validated parameters established at annual validation/performance re-qualification.
- Loads cannot be larger or more difficult to sterilise than the validated “challenge load.”
- Any new load configuration that could exceed the validated challenge load requires re-validation.
- In an emergency situation (where re-validation cannot occur immediately), a Class 5 or 6 chemical indicator may be used until validation can be arranged as soon as possible.
Wrapped and unwrapped items
- Ideally, do not sterilise wrapped and unwrapped items in the same load.
- If unavoidable:
- never place unwrapped items above trays of wrapped items, as condensate can drip onto packs below.
Equipment used when loading the steriliser
- Only use validated, ARTG-listed sterilisation containers, load separators, perforated metal trays, or cassettes.
- Equipment needs to be designed for, and validated with, the specific steriliser model.
- Obtain accessories directly from the steriliser manufacturer or a medical supply company if not supplied with the steriliser.
Load separators are used to space items, for example identical kidney dishes that would stack closely together, hindering air removal, steam penetration and drying.
Trays used in the steriliser need to be:
- perforated to allow steam penetration and drainage of condensation
- specifically validated for the steriliser model.
Multiple trays may be required if packs are sterilised flat in single layers.
Do not use solid trays: they block steam movement, cause pooling of condensate, and may result in wet packs.
Method of loading
- Follow the loading instructions provided by the manufacturer.
- Follow the manufacturer’s instructions on placement of packs and use of equipment.
- Ensure adequate space between each pack/pouch to facilitate air removal, allow steam penetration and facilitate drying – load trays loosely with a single layer of packs on each tray, within the boundaries of the loading tray.
- Load the steriliser so that no items touch the walls. A square frame insert to achieve this may be supplied with the steriliser.
- Loading of the steriliser follows the same pattern of loading described in the validation protocol.
- Photographs or diagrams of the validated loading pattern and any lesser loading patterns used are useful as a reference.
Table 10.17 Principles of steriliser loading
|
DO
|
DON’T:
|
|
Place only a single layer of packs on each tray and allow space between packs for air removal and steam penetration.
Place lighter items on top shelves and heavy items on lower shelves.
For items that are prone to entrap air and condensate, such as bowls, tilt on sides (eg 45° angle using a vertical rack) to allow air removal and steam penetration and facilitate drainage and drying at the completion of the cycle.
Place single laminated pouches vertically in a spacer with paper to laminate, or horizontally, according to steriliser manufacturer’s instructions.
Load items within the boundaries of the shelf.
|
Do not place any items into the steriliser until the complete load is ready to be loaded.
Do not overload the chamber or allow packs to overlap.
Do not allow packages to touch the sides of the steriliser chamber.
Do not exceed the maximum weight or volume specified by the steriliser manufacturer
Do not exceed the challenge load.
Do not place instrument trays, hollow items or items that could collect moisture (eg bowls) above packs containing soft materials.
|